Integral Process Analytical System Control
SP1 Flow Path and Valve Control (Concentration)
The diagram below shows the flow path through the components installed in an SP1. The table below the diagram lists the SP1 valves and the function each valve performs.
Note:
When a solenoid valve is off, port NO is open and port NC is closed. When a solenoid valve is on, port NO is closed and port NC is open.
Valve |
Function |
SP_Solenoid_1 (SV1) |
On—directs stock standard to metering valve (for calibration standard preparation) |
SP_Solenoid_2 (SV2) |
On—directs diluent to dilution pump |
SP_Solenoid_3 (SV3) |
On—directs check standard to sample valve |
SP_Solenoid_4 (SV4) |
On—directs sample or check standard to SS valve and loading pump for analysis. |
SP_Solenoid_5 (SV5) |
On—directs calibration standard to loading pump |
SP_Solenoid_6 (SV6)Dilution Vessel Valve |
On—directs calibration standard to SS valve |
SP_Solenoid_7 (SV7) |
not installed |
SP_Solenoid_8 (SV8) |
On—pressurizes the dilution vessel |
RotaryValve_1 |
Position 10–1—delivers diluent and loop contents to dilution vessel Position 1–2—loads stock standard into the loop for delivery to dilution vessel |